Product features:
Applications:
- Multi-point push-button calibration (up to five points) with ±1% full scale accuracy
- Option of quick, easy automatic calibration or customised, near-to-sample manual calibration
- Auto-ranging for conductivity/TDS measurements for fast response and best resolution over wide measurement range
- Automatic Temperature Compensation (ATC) for optimum accuracy under fluctuating temperatures
- “Hold” function momentarily freezes reading for easy viewing
- Auto power-off saves battery power after non-use
- Self diagnostic with message codes for easy troubleshooting
- Large custom LCD provides optimum viewing even at a distance
- Electrode with built-in ATC designed for minimal air bubble entrapment during measurement
- Protective rubber boot shields meter from drops and features a sturdy built-in stand for easy benchtop operation
- Meter kit (optional) - contains everything you need for calibration and measurement in a rugged carrying case
Applications:
- Routine testing: For quick, accurate conductivity or TDS checks in laboratories, field and schools.
- Environmental/agricultural: Useful in nutrient and fertiliser checks in hydroponics and agricultural industries.
- Water quality testing: For analysing hard water, untreated water, industrial and rinse water, drinking water, effluent water, pool water and incoming process water. Ideal for all types of quality assurance, printing industries and water quality testing
Specifications
Specifications
| Description | Portable conductimeter Fisherbrand accumet AE6C |
| Accuracy (Conductivity) | ±1% Full-scale + 1 LSD |
| Cell Constant | 0.1, 1.0, 10.0 (selectable) |
| Calibration Points (Conductivity) | Four (one per range) auto; five (one per range) manual |
| Range (Conductivity) | 19.9 μS/cm to 199.9 μS/cm, 199.9 μS/cm to 1999 μS/cm, 2.00 mS/cm to 19.99 mS/cm, 20.0 mS/cm to 199.9 mS/cm |
| Resolution (Conductivity) | 0.05% Full-scale |
| Temperature Compensation | Automatic/Manual (0 to 80°C) Coefficient of 0.0% to 3.0% Normalisation of 20.0 to 25.0 |
| Temperature Range (Metric) | 0°C to 80.0°C |
| Resolution (Temperature) | 0.1°C (0.1°F) |
| Accuracy (Temperature) | ±0.5°C (±0.9°F) |