SI Analytics TitroLine™ 5000 Automatic Titrators

Combines a syringe burette and pH/mV meter plus integrated intelligence. SI Analytics TitroLine 5000 Automatic Titrators provide high resolution pH/mV measurement input for pH-, ORP-, silver- and further mV-electrodes.

Overview
Ā£2160.00 valid until 31/12/2016
Log in to get the promotional price



Brand: SI Analytics

Manufacturer Part Number: 285225790

TitroLineĀ® 5000 with 50 ml dosing unit Basic unitwith pH-electrode and buffer set

UNSPSC: 41120000

Code: NEW

Additional Details:
Additional Details: Weight: 7.00000kg



Includes:
Cat. No. 15314408 - TitroLine 5000 with 20mL dosing unit
Ready to use assembled 20mL dosing unit, manual controller, titration clamp, stand rod, magnetic stirrer TM 50 and power supply 100 to 240V

Cat. No. 15324408 - TitroLine 5000 with 50mL dosing unit
pH electrode and buffer set, with ready to use assembled 50mL dosing unit, manual controller, titration clamp, stand rod, magnetic stirrer TM 50 and power supply 100 to 240V

Product Code. 15324408

Quantity Price
1 £ 2711.83 / Each
Estimated Shipment date
from Supplier 06-01-2017
Add to basket


Description and Specification

Specification

Applications Determine chloride in wastewater, permanganate index and COD, total Kjeldhal nitrogen, FOS/TAC, pH value, alkalinity
Includes pH electrode and buffer set, with ready to use assembled 50mL dosing unit, manual controller, titration clamp, stand rod, magnetic stirrer TM 50 and power supply 100 to 240V
Type Automatic Titrators
Connector Types 1 x USB-A and 1 x USB-B, 2 x RS-232-C
Certifications/Compliance ISO 8655, part 6, CE-mark: EMC: 2004/108/EG; safety EG- Directive 2006/95
Depth (Metric) 20.5cm
Display Type Graphics-capable TFT
Hertz 50/60Hz
Temperature Range (Metric) Ambient 10° to 40°C
Width (Metric) 13.5cm
Electrical Requirements 100 to 240V; 50/60Hz, power input 30 VA
Model TitroLine 5000 with 50mL dosing unit

  • pH/mV-input with electrode socket according to DIN 19 262 or additional with BNC socket insert (Z 860)
  • Pt 1000 and NTC 30 temperature measurement input (socket 2 x 4mm) for automatic temperature compensation
  • Pre-installed standard methods for FOS/TAC, alkalinity, total acidity in drinks, chloride
  • Unit operated using the keys on the device itself, the controller TZ 3880 and optional PC-keyboard (USB)
  • Graphics-capable TFT display
  • Linear and dynamic titration to equivalence points
  • Titration to pH and mV-ends points
  • Stirrer Connection: TM 50 power supply directly through piston burette
Specifications
  • Interfaces: 1 x USB-A and 1 x USB-B, 2 x RS-232-C
  • Volume Display: 0000,000 to 9999,999mL
  • Display Resolution: 0.005 to 0.025mL (depending on dosing unit)
  • Filling Speed: min 30 sec. to 999 sec. adjustable (time according to the cylinder volume)
  • Burette Resolution: 8000
  • Dosing Accuracy: systematic error 0.15 %, random error 0.05 % in compliance with EN ISO 8655-6
  • Dosing Speed: max. 100mL/min. (with 50mL unit)
  • Ambient Temperature: 10° to 40°C for operation and storage
  • Humidity: according to EN 61 010, Part 1: Max. relative humidity 80 % for temperatures up to 31°C, linear decrease down to 50 % relative humidity at a temperature of 40°
  • Power Supply: 100 to 240V; 50/60Hz, power input 30 VA
  • Dimensions (D x W x H): 20.5 x 13.5 x 31cm (including dosing unit, without stirrer)
  • Weight: 2.3kg (without stirrer)
Applications
Chloride in wastewater, permanganate index and COD, total Kjeldhal nitrogen, FOS/TAC, pH value, alkalinity

Compliance: ISO 8655, part 6

Certifications: CE-mark: EMC: 2004/108/EG; safety EG- Directive 2006/95

Documentation

Technical documents